| Model Number |
CA2110X12B |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information alleging a trilogy evo "ventilator inoperative condition occurred.The customer states the device turned off while being moved from a roll stand to a wheelchair bracket.The customer states the internal and external batteries were both changed and were both plugged into ac power prior to the device being moved.Once the device was unplugged to be move the device randomly turned off.There was no harm or injury reported.The device was sent to a third-party service center for evaluation.During the evaluation of the device at the third-party service center, error code e-325 was found in the ventilator's downloaded error log.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
Previously reported: the manufacturer received information alleging a trilogy evo "ventilator inoperative condition occurred.The customer states the device turned off while being moved from a roll stand to a wheelchair bracket.The customer states the internal and external batteries were both changed and were both plugged into ac power prior to the device being moved.Once the device was unplugged to be move the device randomly turned off.There was no harm or injury reported.The device was sent to a third-party service center for evaluation.During the evaluation of the device at the third-party service center, error code e-325 was found in the ventilator's downloaded error log.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.New information: correction to section b5: the manufacturer received information alleging a trilogy evo "ventilator inoperative condition occurred.The customer states the device turned off while being moved from a roll stand to a wheelchair bracket.The customer states the internal and external batteries were both changed and were both plugged into ac power prior to the device being moved.Once the device was unplugged to be move the device randomly turned off.There was no harm or injury reported.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported receiving information alleging a trilogy evo "ventilator inoperative condition occurred.The customer states the device turned off while being moved from a roll stand to a wheelchair bracket.The customer states the internal and external batteries were both changed and were both plugged into ac power prior to the device being moved.Once the device was unplugged to be move the device randomly turned off.There was no harm or injury reported.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.The ventilator was evaluated by third party service center and the customer's complaint was not reproduced.The device passes the test.
|
| |
|