• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE Back to Search Results
Model Number M00562340
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a050702 captures the reportable event of snare loop cutting problems.
 
Event or Problem Description
It was reported to boston scientific corporation that a captivator snare was used for polyp removal in the colon during a colonoscopy procedure performed on (b)(6) 2024.During the procedure, the device could not cut with both hot and cold snaring.The procedure was completed with another captivator snare.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a050702 captures the reportable event of snare loop cutting problems.Block h10: one captivator - snares was returned for analysis.The device was carefully inspected, and no problems were found.A functional evaluation noted that it was extended and retracted in the 10- inch loop fixture and the loop could extend and retract well.Besides that, an electrical test was performed, and the device was found within specification.No other device problems were noted.The reported complaint of loop failure to cut was unable to be confirmed since the device returned without any problems on it, and a functional and electrical inspections were performed, and the device was found within specification.It is important to mention that the device cannot be functionally tested by emulating the anatomical/procedural factors encountered during the procedure.The analysis of the available information, product analysis of the returned device and product record review did not identify any evidence of either the alleged issues or any defect on the device.Based on the information available and the returned device analysis, the most probable root cause for the reported complaint is no problem detected since the reported complaint cannot be confirmed.
 
Event or Problem Description
It was reported to boston scientific corporation that a captivator snare was used for polyp removal in the colon during a colonoscopy procedure performed on (b)(6) 2024.During the procedure, the device could not cut with both hot and cold snaring.The procedure was completed with another captivator snare.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18747279
Report Number3005099803-2024-00587
Device Sequence Number6328645
Product Code FDI
UDI-Device Identifier08714729747871
UDI-Public08714729747871
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K202478
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/14/2022
Device Model NumberM00562340
Device Catalogue Number6234
Device Lot Number0024110626
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2024
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date04/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-