• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3507555MC
Medical Device Problem Codes Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 12/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On february 23, 2024, mentor became aware that the right breast implant was removed intact during the explantation procedure.As this report is for the right breast implant, rupture is no longer applicable to this file.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.  as such, the investigation will be closed.  if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.  this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.  this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.  if certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a primary breast augmentation surgery with implantation of a 555cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with bilaterally ruptured breast implants during an in-office visit with a medical professional.Magnetic resonance imaging was also conducted.As a result, the patient underwent bilateral removal and replacement with 450cc mentor memorygel breast implants on (b)(6) 2024.Refer to manufacturing report number 1645337-2024-02079 for the contralateral event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key18750567
Report Number1645337-2024-02080
Device Sequence Number17078833
Product Code FTR
UDI-Device Identifier00081317004257
UDI-Public00081317004257
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Catalogue Number3507555MC
Device Lot Number6777810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/29/2024
Supplement Date Received by Manufacturer02/23/2024
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date03/03/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
-
-