US SURGICAL PUERTO RICO MULTIFIRE ENDO HERNIA; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 174025
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a laparoscopic gastric bypass, a last step to close the mesentery, the stapler would not close properly, and on the second device, the surgeon squeeze the handle it broke, it was still in place but with no resistance.A third device was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant products: 174025, 174025 mf endo hernia 0 4.8 stap x6 (lot#: p3e0339).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: d9 (latest return date), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection of the product noted the instrument had a sulu (single use loading unit) attached with one remaining tack.The handle was damaged and broken off.It was reported that the staple would not close properly.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the instrument handle is not completely cycled affecting the timing of the instrument creating a jammed condition, and then cycling the handle a second time with force causing the handle to disengage.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to fire the stapler, simultaneously press the instrument nose against the tissue, mesh or patch and squeeze the handle as far as it will go, then release.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE ENDO HERNIA
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18750573
Report Number2647580-2024-00865
Device Sequence Number14387760
Product Code GCJ
UDI-Device Identifier10884521070592
UDI-Public10884521070592
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K912097
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number174025
Device Catalogue Number174025
Device Lot NumberP3E0339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/18/2024
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date06/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient SexUnknown
-
-