| Catalog Number |
04Z21-35 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed falsely elevated alinity i stat high sensitivity troponin-i results for a 46-year-old female.The following data was provided (normal range is <14.0 ng/l): sid (b)(6).(b)(6) 2024 initial result 29.5 ng/l, the physician questioned the result.(b)(6) 2024 repeated same sample and the result was <2.7 ng/l.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.Return testing was not performed as returns were not available.Device history record review was performed, which did not show any potential non-conformances, deviations, or non-conformances related to the issue under review.Trending review did not identify any trends.Accuracy testing was completed using panels which mimic patient samples using an in-house retained kit of the complaint lot.All specifications were met indicating that lot 58191ud00 is performing acceptably.Labeling review concludes that the issue is adequately addressed.Based on the investigation, no systemic issue or deficiency of the alinity i stat high sensitivity troponin-i assay was identified.
|
| |
|