ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY I STAT HIGH SENSITIVITY TROPONIN-I; IMMUNOASSAY METHOD, TROPONIN SUBUNIT Back to Search Results
Catalog Number 04Z21-35
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed falsely elevated alinity i stat high sensitivity troponin-i results for a 46-year-old female.The following data was provided (normal range is <14.0 ng/l): sid (b)(6).(b)(6) 2024 initial result 29.5 ng/l, the physician questioned the result.(b)(6) 2024 repeated same sample and the result was <2.7 ng/l.No impact to patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.Return testing was not performed as returns were not available.Device history record review was performed, which did not show any potential non-conformances, deviations, or non-conformances related to the issue under review.Trending review did not identify any trends.Accuracy testing was completed using panels which mimic patient samples using an in-house retained kit of the complaint lot.All specifications were met indicating that lot 58191ud00 is performing acceptably.Labeling review concludes that the issue is adequately addressed.Based on the investigation, no systemic issue or deficiency of the alinity i stat high sensitivity troponin-i assay was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I STAT HIGH SENSITIVITY TROPONIN-I
Common Device Name
IMMUNOASSAY METHOD, TROPONIN SUBUNIT
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18751643
Report Number3005094123-2024-00067
Device Sequence Number19758672
Product Code MMI
UDI-Device Identifier00380740194079
UDI-Public00380740194079
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K202525
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number04Z21-35
Device Lot Number58191UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/18/2024
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date03/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY I PROCESSING MODU, 03R65-01, (B)(6); ALNTY I PROCESSING MODU, 03R65-01, (B)(6)
Patient Age46 YR
Patient SexFemale
-
-