U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO FLAT MARKERS; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO FLAT MARKERS; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSDV0016
Medical Device Problem Codes Nonstandard Device (1420); Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a cori assisted tka surgery, the navio flat markers were flickering and they were unable to be seen at times.The procedure was performed, after a significant delay, with the same device.Patient was not harmed as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: the navio flat markers, part number pfsdv0016, lot number 22mk00040, intended for treatment was returned for evaluation.The reported problem could not be confirmed with a visual inspection.The reported problem was confirmed with a functional evaluation.The flat markers will flicker when placed approximately 1.5 meters from the camera and held approximately perpendicular to the camera.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.A historical review concluded that the lot, serial number or part number reported in this event is related to a corrective/preventive action.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The most likely cause of this event is associated with software settings.As part of corrective action, a software change will update the image settings to optimize flat marker detection.Smith & nephew has initiated a field action to voluntarily correct the cori surgical system.The failure mode will continue to be monitored through complaint investigation and trended through post market surveillance activities.Should any additional information be received the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO FLAT MARKERS
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18751768
Report Number3010266064-2024-00036
Device Sequence Number12332425
Product Code OLO
UDI-Device Identifier00885556628904
UDI-Public00885556628904
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Repair,Recall,Notification
Type of Report Initial,Followup
Report Date (Section B) 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/08/2024
Device Catalogue NumberPFSDV0016
Device Lot Number22MK00040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2024
Supplement Date Received by Manufacturer05/09/2024
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date05/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
(B)(6)/REAL INTELLIGENCE CORI
Patient SexUnknown
-
-