| Catalog Number |
8713032U |
| Medical Device Problem Codes |
Excess Flow or Over-Infusion (1311); Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: propofol pump not dosing properly and blood backing up into the tubing.Pharmacist tried to reprogram the pump and it would not work.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun mecical inc.Internal report (b)(4).The actual device involved in the reported incident was returned to b.Braun medical for evaluation and the logs were provided for review.Investigation results: the reported issue was not present during investigation.Volumetrics infusion of 100ml/hr and 10ml (dl: prop10) tested in spec at 9.9ml or 99% of e.Performed preventive maintenance service.Log review shows on (b)(6) 2024 and 12:59pm an infusion start of 42ml/hr and 60ml (dl: prop10).At 1:03pm infusion was stopped with a volume infused to 2.84ml.At 1:21pm with a patient weight change from 70kg to 80kg an infusion start of 67.2ml/hr.At 1:25pm with a volume infused to 7.32ml infusion was stopped with no further infusions taking place.All information concerning this reported incident has been included in our trend analysis of the product line.
|
| |
|