| Model Number |
0998-UC-3013-53 |
| Medical Device Problem Code |
Device Emits Odor (1425)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cs300 intra-aortic balloon pump (iabp) smelled when tank was opened.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the cs300 intra-aortic balloon pump (iabp) smelled when tank was opened.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5, h6 (clinical & impact code).Updated data: b4, g3, g6, h2, h10, h11.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected fields: e3, h6 (health effect ¿ clinical code) it was reported that the cs300 intra-aortic balloon pump (iabp) had odor related malfunction.The patient involvement is unknown but no harm reported.Customer reports smell from tank.Getinge field service engineer (fse) arrived onsite and was not able to verify smell from helium tank however there was a odor coming from the battery compartment of the unit.Replaced batteries due to swollen hot batteries.Completed repair, calibration, full functional tests and safety tests as per the service manual.Unit passed all tests.Returned unit to customer for use.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only** providing updated device identification information in alignment with gudid.
|
| |
|