• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-UC-3013-53
Medical Device Problem Code Device Emits Odor (1425)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cs300 intra-aortic balloon pump (iabp) smelled when tank was opened.
 
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event or Problem Description
It was reported that the cs300 intra-aortic balloon pump (iabp) smelled when tank was opened.There was no patient harm.
 
Additional Manufacturer Narrative
Corrected data: b5, h6 (clinical & impact code).Updated data: b4, g3, g6, h2, h10, h11.
 
Additional Manufacturer Narrative
Corrected fields: e3, h6 (health effect ¿ clinical code) it was reported that the cs300 intra-aortic balloon pump (iabp) had odor related malfunction.The patient involvement is unknown but no harm reported.Customer reports smell from tank.Getinge field service engineer (fse) arrived onsite and was not able to verify smell from helium tank however there was a odor coming from the battery compartment of the unit.Replaced batteries due to swollen hot batteries.Completed repair, calibration, full functional tests and safety tests as per the service manual.Unit passed all tests.Returned unit to customer for use.
 
Event or Problem Description
N/a.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
**udi related data quality updates only** providing updated device identification information in alignment with gudid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
henry pena
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key18754337
Report Number2249723-2024-00765
Device Sequence Number14387825
Product Code DSP
UDI-Device Identifier10607567107745
UDI-Public(01)10607567107745
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K031636
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-UC-3013-53
Device Catalogue Number0998-UC-3013-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2024
Supplement Date Received by ManufacturerNot provided
05/17/2024
01/22/2025
Initial Report FDA Received Date02/21/2024
Supplement Report FDA Received Date03/28/2024
05/20/2024
01/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2004
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient SexUnknown
-
-