• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: I-SENS RELION PREMIER COMPACT BLOOD GLUCOSE TEST SYSTEM; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

I-SENS RELION PREMIER COMPACT BLOOD GLUCOSE TEST SYSTEM; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number GM01XBB
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the complaint meter was not returned, i-sens analyzed the accuracy issue with a reference meter and retained strips from the same lot.Firstly, as a result of the control solution test, all results satisfied the acceptable range in both a and b levels.Secondly, we conducted the clinical test and all 30 data satisfied the accuracy standard of iso 15197:2013.Therefore, we confirmed that the product is functioning properly.
 
Event or Problem Description
F069321k6470.Lot ul12bbehe - 2024-08-11 - 100ct.The patient is receiving high readings.The patient has had meter for about three months and has just recently had an issue with it.The patient was taken to the hospital because of the readings, and they did a blood draw.The patient has not done a control solution test and walmart did not offer control solution and caller was not aware.The caller is contacting us on behalf of the patient.The patient keeps strips in the original bottle.The patient keeps meter and test strips in bedroom away from humidity and direct sunlight.The last five readings are 301, 427 at 6 am, 245 last night, 354 during day yesterday, and 422 yesterday sometime.This meter does not record the date and times of the exact readings, so i was not able to get them.The patient is not on oxygen therapy.The patient keeps strips in the original bottle and the lid closed at all times.The patient has no difficulty getting a blood sample.The patient changes lancets everyday but not always before each test.The patient tests on fingertips only and uses an alcohol wipe before each test.The patient uses the second drop of blood.Today the reading was in the 400's and they were concerned about it, so the caller took the patient to the hospital, and they took his blood and said that his sugar was not in the 400's.The patient was not experiencing any symptoms at the time.The hospital used a fingerstick and it was around 180 and then did a blood draw later and came back around 180.About 30 minutes had passed between relion compact meter reading and hospital reading.The patient only does insulin usually once per week but did take metformin based off the reading this morning of 427 and then went to the hospital.Replaced meter, strips, sent control solution and return label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELION PREMIER COMPACT BLOOD GLUCOSE TEST SYSTEM
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
I-SENS
43, banpo-daero 28-gil
seocho-gu
seoul, 06646
KS  06646
Manufacturer (Section G)
I-SENS WONJU FACTORY
94-1, donghwagongdan-ro
munmak-eup
wonju-si, 26365
KS   26365
Manufacturer Contact
hyojin baik
43, banpo-daero 28-gil
seocho-gu
seoul, 06646
KS   06646
MDR Report Key18754862
Report Number3005621222-2024-00002
Device Sequence Number11688192
Product Code NBW
UDI-Device Identifier00681131169776
UDI-Public(01)30681131240728,(01)00681131169776
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120759
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/11/2024
Device Model NumberGM01XBB
Device Catalogue Number783110
Device Lot NumberUL12BBEHE
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/24/2024
Initial Report FDA Received Date02/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/18/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age42 YR
Patient SexMale
Patient Weight113 KG
-
-