One opened probe was received, with a tip protector, in a tray, for the report.The sample was visually inspected and found to be nonconforming with the needle slightly bent.The sample was then functionally tested for actuation and cut.The sample was found to be conforming for cut and nonconforming for actuation.The probe was disassembled, and the components inspected.Nominal wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard pictures.Gouge marks observed at locations along the inner cutter.The inner cutter was observed to be bent.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The complaint evaluation does not confirm the probe had a cut failure; the evaluation indicates the probe had an actuation failure.The cut functionality of the probe was conforming; however, the observed actuation failure could have caused the probe to not cut at the time the reported event was observed.The most likely cause for the actuation failure is from the bent needle and bent inner cutter.A bent needle can impede the movement of the cutter shaft.This interference is present as observed from visual condition of the inner cutter.The exact root cause of the bent needle and bent inner cutter cannot be determined from this evaluation.The most likely root cause is handling at any point after the probe was shipped from the manufacturing site including use during surgery.The reported cut failure was not confirmed, and an exact root cause for the bent needle and bent inner cutter could not be determined from this evaluation; therefore, no specific action with regard to this complaint was taken by the manufacturing site.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|