Upon troubleshooting it was determined the power cord needed to be replaced.Physical damage of a cord would be readily noticeable to the user.The user would notice any integrity issues with the surrounding insultation (outer covering) when plugging or unplugging the device into the electrical outlet.If found to be damaged, the power cord and device would be immediately removed from use.The power supply cords, by design, meet (iec 60601-1-6 usability standard), (iec 60245-1:2003, annex a, designation 53) or (iec 60227-1:1993, annex a, designation 53).Additionally, specific electrical safety standards apply to each product and subsequent applicable usability testing.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this device.A power cord replacement was shipped direct to patient to resolve the reported event.Based on this information, no further action is required.
|