• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL G2 CURVED JAW 35CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL G2 CURVED JAW 35CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number NSLG2C35
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 2/22/2024.D4 batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a lap sigma the instrument has no longer released.No patient harm nor significant delay reported.Procedure finished with same like device.
 
Additional Manufacturer Narrative
(b)(4), date sent: 4/16/2024.D4 batch #: a9dy9h.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with the top of the i-blade lower tip broken off and not returned.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.Please reference the instruction for use for more information: as part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch a9dy9h, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 CURVED JAW 35CM
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18758552
Report Number3005075853-2024-01531
Device Sequence Number6601773
Product Code GEI
UDI-Device Identifier10705036014140
UDI-Public10705036014140
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNSLG2C35
Device Lot NumberA9DZ91
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/05/2024
Supplement Date Received by Manufacturer04/01/2024
Initial Report FDA Received Date02/22/2024
Supplement Report FDA Received Date04/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-