| Model Number |
CNWTT0 |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implant procedure, when lens was inserted a scratch was noted.The lens cut out and replaced during the initial procedure.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The account indicated the use of a qualified associated cartridge with a company handpiece.Associated viscoelastic was not provided.It is unknown if qualified product was used.The product investigation could not identify a root cause for the reported complaint.The account indicated the product was not available to evaluate.Per the information provided a company handpiece was used with the company cartridge.The company handpiece will not physically accommodate the company cartridge.This may have been reported in error.Associated viscoelastic was not provided.It is unknown if qualified product was used.The instructions for use (ifu) instructs: company intra ocular lens (iol) are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.Follow up attempts were made.No further information has been provided at this time.File will be reopened when new information is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|