| Model Number |
MEE-2000A |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that an electric a stim box #732 did not work on a case.The customer changed both the cable and stim box from another machine and the system worked.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported that an electric a stim box #732 did not work on a case.The customer changed both the cable and stim box from another machine and the system worked.At the end of the case, the customer tried another protocol with stim box 732 and it did not work again.The stim box is being exchanged.There was no patient injury reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.D10 concomitant medical device: the following device was used in conjunction with the mee-2000: js-201b: model #: js-201b.Serial #: (b)(6).Device manufacturer date: 03/24/2021.Unique identifier (udi) #: (b)(4).Returned to nihon kohden: no.
|
| |
|
Event or Problem Description
|
|
The customer reported that an electric a stim box # (b)(6) did not work on a case.The customer changed both the cable and stim box from another machine and the system worked.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: the customer reported that an electric a stim box # (b)(6) did not work on a case.The customer changed both the cable and stim box from another machine and the system worked.At the end of the case, the customer tried another protocol with stim box (b)(6) and it did not work again.The stim box is being exchanged.There was no patient injury reported.Investigation summary: multiple follow-up requests for the return of the complaint device were sent to the customer but they were unresponsive.A definitive root cause could not be determined as we have not received the complaint device for evaluation.The following fields contain no information (ni), as attempts to obtain information were made, but not provided: a2 - a6 attempt # 1: on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.Attempt # 2: on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.Attempt # 3 on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.B6 attempt # 1: on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.Attempt # 2: on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.Attempt # 3 on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.B7 attempt # 1: on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.Attempt # 2: on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.Attempt # 3 on (b)(6) 2024 emailed the customer via microsoft outlook for patient information: no reply was received.D4 serial number attempt # 1: on (b)(6) 2024 emailed the customer via microsoft outlook for device information: no reply was received.Attempt # 2: on (b)(6) 2024 emailed the customer via microsoft outlook for device information: no reply was received.Attempt # 3 on (b)(6) 2024 emailed the customer via microsoft outlook for device information: no reply was received.The following fields are not available (n/a) to this report: h4 device manufacturer date; d10 concomitant medical device: the following device was used in conjunction with the (b)(6).: js-201b: model #: js-201b serial #: (b)(6).Device manufacturer date: (b)(6) 2021 unique identifier (udi) #: (b)(4).Returned to nihon kohden: no.Additional information: b4 date of this report g3 date received by manufacturer g6 type of report h2 if follow up, what type? h11 additional manufacturer narrative: manufacturer references # (b)(4).Follow up 001.
|
| |
|