| Catalog Number |
1012448-08 |
| Medical Device Problem Codes |
Inflation Problem (1310); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaints.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat an 80% stenosed lesion in the proximal circumflex (cx) artery with heavy calcification and moderate tortuosity.The protective sheath of the 2.75x8mm nc trek balloon dilatation catheter (bdc) was removed with difficulty.After pre-dilatation and implantation of a non-abbott stent, the nc trek bdc was advanced for post-dilatation; however, the balloon failed to inflate.Several attempts were made to inflate the balloon and the inflation device was replaced; however, the balloon still did not inflate.Therefore, the bdc was removed from the patient and another same size nc trek was used to complete the procedure.There was no adverse patient effect and no clinically significantly delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|