• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MOSAIC MITRAL BIOPROSTHETIC HEART VALVE; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MOSAIC MITRAL BIOPROSTHETIC HEART VALVE; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 310C33
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Non specific EKG/ECG Changes (1817)
Date of Event 01/22/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Medtronic received information that during the implant of this 36mm mitral annuloplasty ring, it was explanted and replaced with a 33mm mitral bioprosthetic valve.The reason for the replacement was reported as moderate regurgitation.It was noted that the mitral valve had demonstrated severe mitral annular dilation with a large flail p3 segment.It was stated that the p3 segment, which was flail of the posterior mitral leaflet, was resuspended the ring was then implanted and sutures were adjusted to minimize the degree of mitral insufficiency and a posterior medial commissuroplasty was performed.It was reported that inspection of the valve with saline infusion demonstrated continued ongoing probably at least moderate regurgitation present near the hood of the p3 prolapsed segment.An attempt was made to minimize the degree of mitral regurgitation by tucking the segment down to the p2 segment, however, this was unsuccessful and it was decided to replace the valve.It was reported that the patient received one bout of defibrillation fol lowing the valve replacement and then returned to spontaneous cardiac contraction.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Product analysis: the device remains implanted, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Conduction disturbances are known potential adverse effects associated with any cardiac or thoracic procedure (open or catheter-based) and can be resolved with medical treatment or the implant of a permanent pacemaker (with the risk-benefit ratio in favor of implant of the percutaneous aortic valve).A conduction disturbance does not indicate a device malfunction or potential manufacturing issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC MITRAL BIOPROSTHETIC HEART VALVE
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18760712
Report Number2025587-2024-01070
Device Sequence Number12688268
Product Code DYE
UDI-Device Identifier00643169001572
UDI-Public00643169001572
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P990064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number310C33
Device Catalogue Number310C33
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2024
Initial Report FDA Received Date02/22/2024
Date Device Manufactured11/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight90 KG
Patient RaceWhite
-
-