| Model Number |
VTICM5_13.2 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Claim #: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted an implantable collamer lens into the patient's left eye (os).It was indicated there was rotation without low vaulting.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h6- health effect - impact code (f): "2199" should be removed and "4629" and "4628" should be added.H6- medical device problem code (a): "2993" should be added.H6- type of investigation code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.B5- the reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens of diopter -13.00/3.0/103 (sphere/cylinder/axis) into the patient's left eye (os) (b)(6) 2023.The patient experienced a lens rotation not associated with a low vault.On (b)(6) 2023 the lens was repositioned but that did not resolve the problem.On (b)(6) 2024 the lens was exchanged with a different model, longer length lens and the problem was resolved.The status of the eye was noted to be "all good, patient is happy".The reporter indicated the cause of the event is unknown.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h3 - type of investigation code: 10- device evaluation: the lens was returned dry in a lens case.Visual inspection found optic deformed.Haptic bent/ stretched out.Dimensional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|