| Model Number |
EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 11MM |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
01/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products equinoxe reverse 42mm humeral liner +0 (cat# 320-42-00 / serial# (b)(6)) equinoxe reverse tray adapter plate tray +0 (cat# 320-10-00 / serial# (b)(6)) eq reverse torque defining screw kit (cat# 320-20-00 / serial# (b)(6)) additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, approximately three months post initial right tsa, the 58 y/o male patient presented with recurring dislocation due to undersized/loose stem.Patient had a replacement of stem, liner, glenosphere, torque screw, and locking screw.There was no reported breakage of device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Sales rep was unable to obtain images or x-rays.The devices are not available for evaluation due to they were disposed by hospital.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was likely the result of humeral stem loosening and subsequent dislocation.The loosening may have been due to undersizing the stem, as stated in the experience report.However, this cannot be confirmed because the component was not returned for evaluation and radiographs were not provided.
|
| |
|