| Catalog Number |
33680001 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Inadequate Osseointegration (2646); Limb Fracture (4518)
|
| Date of Event |
01/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.H3 other text : device remains implanted in patient.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a removal surgery of the current talar implant due to early collapse of the tibial side without trauma.There is a fracture present in the tibia.The physician reported that the patient will need intramedullary fixation.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a removal surgery of the current talar implant due to early collapse of the tibial side without trauma.There is a fracture present in the tibia.The physician reported that the patient will need intramedullary fixation.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed, confirmed based on available ct scans and medical expert assessment.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed: "subsidence and loosening of the tibial tray due to poor bone quality and a fracture of the tibial plafond, thus patient related.No early postoperative x-ray for comparison.Components as such intact".Based on the investigation, root cause was attributed to patient related issue.The failure was caused due to the poor bone quality along with fracture of tibial plafond observed in the ct scans.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|