| Model Number |
64082 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Premature Separation (4045)
|
| Health Effect - Clinical Code |
Laceration(s) of Esophagus (2398)
|
| Date of Event |
02/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
According to the reporter, the patient had a laceration when the catheter was removed.According to medical staff, an esophageal stent was placed on patient.Patient was noted to be fine and in recovery.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the patient had a laceration when the catheter was removed after ablation procedure and upon examination, the balloon catheter broke or was not intact.The scope was advanced into the esophagus to find the piece of balloon and it was still in the esophagus.The piece of balloon was removed.Superficial mucosal involvement, approximately 5 cm, but with blood and mucus exact measurement not possible, but appeared superficial.According to medical staff, an esophageal stent was placed on patient.An x-ray and barium swallow was performed after the removal of the stent.Patient stayed a couple of days due to initial nausea.Patient was noted to be fine and in recovery.
|
| |
|
Search Alerts/Recalls
|