• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INFINITY ADAPTIS TALDOME SZ1 INFINITY ADAPTIS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INFINITY ADAPTIS TALDOME SZ1 INFINITY ADAPTIS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 33680021
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Limb Fracture (4518)
Date of Event 01/25/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.H3 other text : device remains implanted in patient.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a removal surgery of the current talar implant due to early collapse of the tibial side without trauma.There is a fracture present in the tibia.The physician reported that the patient will need intramedullary fixation.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, confirmed based on available ct scans and medical expert assessment.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Upon further investigation of the ct scans by healthcare professionals the following was observed: "subsidence and loosening of the tibial tray due to poor bone quality and a fracture of the tibial plafond, thus patient related.No early postoperative x-ray for comparison.Components as such intact".More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a removal surgery of the current talar implant due to early collapse of the tibial side without trauma.There is a fracture present in the tibia.The physician reported that the patient will need intramedullary fixation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY ADAPTIS TALDOME SZ1 INFINITY ADAPTIS
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18764013
Report Number3010667733-2024-00094
Device Sequence Number4387795
Product Code HSN
UDI-Device Identifier00889797101516
UDI-Public00889797101516
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K172633
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number33680021
Device Lot Number1735733
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2024
Supplement Date Received by Manufacturer04/22/2024
Initial Report FDA Received Date02/22/2024
Supplement Report FDA Received Date05/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-