Lot Number 0032134034
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the moderately tortuous and moderately calcified coronary artery.A 3.50 x 32mm synergy megatron drug-eluting stent was advanced for treatment.However, during the procedure, the stent catheter was broken.The device was removed completely intact from the patient, and the procedure was completed with another of the same device.There were no patient complications reported.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the moderately tortuous and moderately calcified coronary artery.A 3.50 x 32mm synergy megatron drug-eluting stent was advanced for treatment.However, during the procedure, the stent catheter was broken.The device was removed completely intact from the patient, and the procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: synergy megatron mr us 3.50 x 32mm was returned to the complaint investigation site (cis).A visual, tactile and microscopic examination of the hypotube shaft identified a break in the device at 25.8cm distal to the distal end of the strain relief.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues.There was no sign of damage, stretching or lifting of the stent struts.Balloon cones were reviewed, and no issues were noted.Balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.No signs of movement, stent was set between the proximal and distal markerbands.Bumper tip showed signs of distal tip damage (flared).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY MEGATRON
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18764021
Report Number2124215-2024-10550
Device Sequence Number11887043
Product Code NIQ
UDI-Device Identifier08714729985778
UDI-Public08714729985778
Combination Product (Y/N)Y
Initial Reporter StatePR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/16/2025
Device Lot Number0032134034
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2024
Supplement Date Received by Manufacturer03/26/2024
Initial Report FDA Received Date02/22/2024
Supplement Report FDA Received Date03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
-
-