• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY XR2 SKULL CLAMP Back to Search Results
Catalog Number A2114
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 01/30/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mayfield infinity xr2 skull clamp (a2114) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned unit showed that the unit had both the right and left pawls broken.The unit is over three years old and has no history of service.The 80lb torque knob is sticking, the index knob is loose and due to lack of service, it is recommended that the unit have a preventative maintenance (pm) service.Additionally, the unit was received without the black case or the pediatric rocker arm.Since the unit was involved in a patient injury, it was sent to quality engineering (qe) for further investigation.Qe confirmed the findings of the service & repair report.New components were added to replace worn internal parts, and pm performed.Root cause - complaint confirmed via inspection.The unit has no service history and requires preventive maintenance and replacement of worn parts.Additionally, improper, or suboptimal placement of the skull clamp can contribute to movement of the patient¿s head.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Event or Problem Description
A facility reported that during spine surgery, a staff member was transferring the patient pinned in the mayfield infinity xr2 skull clamp (a2114) from the transport bed and when they attempted to turn the patient from supine to prone position, the plunger disengaged.This caused the patient's head to slip out of the pins resulting in a minor superficial laceration that did not require stitches.There were slight changes in the patient's vital signs as per neuro monitoring following the incident, prompting the anesthesiologist to request to abort the case.No surgical delay have been reported.Additional information received as follows: anesthesia ask to abort the case".Was the procedure cancelled, or was it rescheduled to another date? "i do not have any further information.All the information written on the paper is all i got.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY XR2 SKULL CLAMP
Common Device Name
MAYFIELD
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18764502
Report Number3004608878-2024-00022
Device Sequence Number7739322
Product Code HBL
UDI-Device Identifier10381780253761
UDI-Public10381780253761
Combination Product (Y/N)N
PMA/510(K) Number
K130389
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberA2114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2024
Initial Report FDA Received Date02/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-