Model Number CC60WF
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The used lens appeared to be replaced into the associated company cartridge for return shipment inside the lens carton.Viscoelastic was observed on the lens.The lens was placed posterior surface up and located halfway into the loading area.The leading haptic was extended.The trailing haptic was folded onto the anterior surface of the optic and adhered in dried viscoelastic.There was no damage observed to the associated cartridge.Both wings display evidence of handpiece use.Product history records were reviewed, and documentation indicated the product met release criteria.All listed associated products were qualified for use with the lens.The root cause cannot be determined for the reported complaint of "didn't eject properly, in and out".It is unknown what the customer meant by "didn't eject properly".The lens was returned replaced halfway into the loading area of the associated used cartridge.No damage was observed to the cartridge.Due to the method of return, no further determination can be made.+ the completed questionnaire indicated that in the surgeon's opinion, the product did not cause or contribute to the event.The root cause per the surgeon was unknown.Information was provided that the company lens was removed and replaced with the same model lens.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-health care professional reported that during an intraocular lens (iol) implant procedure,lens did not eject properly.Lens was explanted at initial procedure.Additional information was received.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18765514
Report Number1119421-2024-00332
Device Sequence Number18077424
Product Code HQL
UDI-Device Identifier00380652396790
UDI-Public00380652396790
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCC60WF
Device Lot Number15706826
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2024
Initial Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR.; MONARCH II LOADING FORCEPS.; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.; VISCOAT OPHTHALMIC VISCOSURGICAL DEVICE.
Patient Age68 YR
Patient SexFemale
-
-