| Catalog Number |
VENUM18120 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiration date: 08/2024).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure in the iliac vein, the proximal end of the stent allegedly did not fully detach from the delivery system.It was further reported that even after one minute, the stent allegedly did not release.Reportedly the physician rotated the deployment handle and completed the procedure.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure in the iliac vein, the proximal end of the stent allegedly did not fully detach from the delivery system.It was further reported that even after one minute, the stent allegedly did not release.Reportedly the physician rotated the deployment handle and completed the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: a voluntary recall has been initiated for the venovo¿ venous stent system 9f which was product catalog/lot number specific.Reportedly the proximal end of the stent does not immediately expand upon deployment.The proximal end of the stent remains connected to the delivery system.As a result of the field action, this event is being reported as a malfunction reportable event.H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation.No images or videos were provided for review.Based on the information available the investigation is closed with inconclusive result for stent expansion issue.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant ifu for this product was conducted.The ifu was found to address potential worst case complications and adverse events potentially related to expansion issue such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, stent fracture, and malposition.H10: d4 (expiration date: 08/2024), g3 h11: h6 (method) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|