Catalog Number PDP339H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a total laparoscopic hysterectomy procedure on (b)(6) 2024 and suture was used.During the procedure, it was reported by a sales rep that at the time of suture closure of vaginal stump, the needle was detached from the suture at the 2nd stitch.Another package was opened and used but the same thing happened again.These were not control release needles.Suture method was continuous suture.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, and the following was obtained via: "it was reported that: ""at the time of suture closure of vaginal stump, the needle was detached from the suture at the 2nd stitch.Another package was opened and used but the same thing happened again"".* according to event description, it appears (b)(4) devices were affected, however, quantity of product involved was entered as (b)(4).Please confirm the number of devices affected during this procedure.= (b)(4).* please confirm the quantity of devices returning for analysis.= (b)(4).* two lots were provided: sdmqtq, sdmrdk.Please confirm the number of sutures affected per lot.= it could not be determined.* please provide the source or name and title of the external person providing answers to follow-up (external person submitting answers to sales rep)= (b)(6).Please perform and document the follow-up attempt for product return.Please document the shipment tracking number in the notes or rmao sections.= we regularly contact with sales rep about the device returning.Events reported via; (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDS+ VIO 27IN 2-0 S/A CT-1
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key18767007
Report Number2210968-2024-01723
Device Sequence Number12711522
Product Code NEW
UDI-Device Identifier10705031048096
UDI-Public10705031048096
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K061037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Catalogue NumberPDP339H
Device Lot NumberSDMRDK
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2024
Initial Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-