MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PIN PASSING DRILL TIP 2.4X381MM STRL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PIN PASSING DRILL TIP 2.4X381MM STRL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 7208678
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopy, at the time of passing through the femur for the tunnel, the low profile drill bit was inserted to tunnel the graft, and it was noticed that the pin passing drill tip was worn, when the drill bit passed through it to make the tunnel and began to release particles of the pin material leaving the material dispersed in the patient's joint.It was reported that the patient current status is stable.The procedure was completed with non-significant delay using a back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
H6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Event or Problem Description
It was reported that during an arthroscopy, at the time of passing through the femur for the tunnel, the low-profile drill bit was inserted to tunnel the graft, and it was noticed that the pin passing drill tip was worn, when the drill bit passed through it to make the tunnel and began to release particles of the pin material leaving the material dispersed in the patient's joint.The remaining particles were attempted to be recovered using a shaver, since they were very small and only by using a blade and suction it was possible to obtain a large quantity of them.Some remained because due to the material of the pin they adhered to the tissues, but they were very small particles.It was reported that the patient current status is stable.The procedure was completed with non-significant delay using a back-up device in the originally drilled bone hole and no void was left in the patient.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).H10 b1, b2, b5, h1 and h6: event description, type of reportable event and codes updated.
 
Additional Manufacturer Narrative
H10: h3, h6: a device deficiency was identified, the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that one undated image of the device was provided for review and confirms missing coating from the instrument; however, it does not indicate a root cause.All documents and images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical/medical investigation.Based on the limited information provided the clinical root cause of the reported events could not be determined and we are currently unable to rule out a procedural variance as a contributing factor to the reported event, which does not represent a device malfunction.The material of the pin passing drill is intended for surgical applications and is biologically compatible; however, this device is not approved for implantation; therefore, long-term implantation data is not available.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
 
Additional Manufacturer Narrative
H10 h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection of the returned device finds the drill has some small scratches and biological debris on the distal tip, the shaft is abraded and the surface lifted in areas along the distal half.A clinical review states that one undated image of the device was provided for review and confirms missing coating from the instrument; however, it does not indicate a root cause.All documents and images provided as of this date have been reviewed and considered and unless noted do not contribute to the clinical/medical investigation.Based on the limited information provided the clinical root cause of the reported events could not be determined and we are currently unable to rule out a procedural variance as a contributing factor to the reported event, which does not represent a device malfunction.The material of the pin passing drill is intended for surgical applications and is biologically compatible; however, this device is not approved for implantation; therefore, long-term implantation data is not available.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN PASSING DRILL TIP 2.4X381MM STRL
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18767748
Report Number1219602-2024-00336
Device Sequence Number12674028
Product Code HWE
UDI-Device Identifier03596010591265
UDI-Public03596010591265
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number7208678
Device Lot Number2130006
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2024
Supplement Date Received by Manufacturer04/04/2024
04/08/2024
05/14/2024
06/25/2024
Initial Report FDA Received Date02/23/2024
Supplement Report FDA Received Date04/05/2024
04/18/2024
05/16/2024
06/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexMale
Patient Weight90 KG
-
-