| Model Number |
24700 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Thrombosis/Thrombus (4440)
|
| Date of Event |
02/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
B3: date of event - used the first date of the month of the aware date as no event date was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon loading tool was inadvertently left in the patient.The patient presented with a thrombosed arteriovenous fistula (avf) and underwent thrombolysis and angioplasty.Ultrasound revealed that a long and tubular foreign body was in the avf from a previous procedure.The foreign body was believed to be a balloon cover of a 4.0mmx100mmx80cm sterling balloon catheter which was accidentally pushed into the introducer sheath and into the blood vessel.A guidewire and balloon were used to push the foreign body out of the vessel and treat the thrombosed avf.The procedure was completed, and the patient was fully recovered.
|
| |
|