• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Thrombosis/Thrombus (4440)
Date of Event 02/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B3: date of event - used the first date of the month of the aware date as no event date was provided.
 
Event or Problem Description
It was reported that the balloon loading tool was inadvertently left in the patient.The patient presented with a thrombosed arteriovenous fistula (avf) and underwent thrombolysis and angioplasty.Ultrasound revealed that a long and tubular foreign body was in the avf from a previous procedure.The foreign body was believed to be a balloon cover of a 4.0mmx100mmx80cm sterling balloon catheter which was accidentally pushed into the introducer sheath and into the blood vessel.A guidewire and balloon were used to push the foreign body out of the vessel and treat the thrombosed avf.The procedure was completed, and the patient was fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18768643
Report Number2124215-2024-09993
Device Sequence Number11906915
Product Code DQY
UDI-Device Identifier08714729123514
UDI-Public08714729123514
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24700
Device Catalogue Number24700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2024
Initial Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-