| Catalog Number |
06473245001 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The gluc3 reagent lot number and expiration date were not provided.Qc was acceptable on the day of the event.On 19-jan-2024, preventive maintenance was performed and the gear pump head was replaced.A general reagent issue can be excluded as a correct rerun was performed with the same reagent.The field service engineer (fse) found 2 leaky vacuum tubes in the cell rinse.He replaced the vacuum tubes.The fse also replaced the degasser and adjusted the gear pump pressure value.The fse did a performance check and the instrument was performing within the specifications.The cause was related to a maintenance issue.The service actions (replacing the vacuum tubes in the cell rinse) resolved the issue.
|
| |
|
Event or Problem Description
|
|
We received an allegation about discrepant results for an unspecified number of patient samples tested with different assays on a cobas 8000 c702 analyzer.Discrepant results for 1 patient serum sample tested with glucose (gluc3) assay were provided.Initial result: 28.2 mg/dl.Repeat result: 65.7 mg/dl.The repeat result was deemed to be correct.
|
| |
|
Search Alerts/Recalls
|