Catalog Number 4566
Medical Device Problem Codes Break (1069); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during an arthroscopic rotator cuff repair, the regeneten bioinductive metal implant arms dislodged from deployment gun inside of shoulder joint.It was successfully retrieved from the patient.The procedure was successfully completed using the reported device.There was a surgical delay less than 30 minutes and no further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).H10 h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found that the cartridge head top is broken, and the ds finger fell off from the device.The broken cartridge head top has been determined to be related to the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause of the reported event could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include an application of inappropriate or excessive force on the device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOINDUCTIVE IMPLANT W/ARTH DEL LRG
Common Device Name
MESH, SURGICAL, COLLAGEN, ORTHOPAEDICS, REINFORCEMENT OF TENDON
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18768770
Report Number3003604053-2024-00007
Device Sequence Number20744701
Product Code OWY
UDI-Device Identifier00885556735664
UDI-Public885556735664
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K140300
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2026
Device Catalogue Number4566
Device Lot Number2123446
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2024
Supplement Date Received by Manufacturer01/30/2024
Initial Report FDA Received Date02/23/2024
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-