• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ7 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ7 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 157012135
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Synovitis (2094); Osteolysis (2377); Metal Related Pathology (4530); Swelling/ Edema (4577)
Date of Event 11/13/2009
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).E3 initial reporter occupation: lawyer.H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Update ad 15 feb 2024 medical records received.Patient underwent revision due to elevated cobalt levels- removal of cobalt chrome hip ball and cobalt chrome acetabular liner.Slightly cloudy yellow synovial fluid in moderate quantity, associated with reactive, tan-colored synovitis consistent with metallosis, trunnion corrosion identified.This included small thin pieces of dark stained material adherent to the base of the trunnion.Periprosthetic osteolysis and wear of articular bearing surface.For placement with highly cross-linked polyethylene liner and revision of ceramic hip ball.Doi: (b)(6) 2009, dor: (b)(6) 2021.Affected side : left hip.This is a bilateral hip arthroplasty.A link of (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ7 HI OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18769352
Report Number1818910-2024-04205
Device Sequence Number20782466
Product Code LPH
UDI-Device Identifier10603295060185
UDI-Public10603295060185
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K193398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/14/2019
Device Catalogue Number157012135
Device Lot NumberD32E21
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2024
Initial Report FDA Received Date02/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/16/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12/14 ARTICUL 40MM M SPEC-2; APEX HOLE ELIM POSITIVE STOP; PINN CAN BONE SCREW 6.5MMX20MM; PINN CAN BONE SCREW 6.5MMX25MM; PINN CAN BONE SCREW 6.5MMX25MM; PINNACLE MTL INS NEUT40IDX56OD; PINNACLE SECTOR II CUP 56MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexMale
-
-