Model Number N/A
Medical Device Problem Codes Break (1069); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/31/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will evaluate.Results are expected soon.
 
Event or Problem Description
It was reported, "placed a midline catheter on a patient in the or.Used the powerglide as is, however, the catheter broke and a surgeon had to come and do a cut down to get the retained wire/midline catheter.To their best understanding, the catheter broke off, but the surgeon was able to remove successfully." no other information was provided.
 
Event or Problem Description
It was reported, "placed a midline catheter on a patient in the or.Used the powerglide as is, however, the catheter broke and a surgeon had to come and do a cut down to get the retained wire/midline catheter.To their best understanding, the catheter broke off, but the surgeon was able to remove successfully." no other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a broken catheter is confirmed and was determined to be use related.One 22 ga powerglide pro was returned for evaluation.An initial visual observation showed blood use residues throughout the returned powerglide pro.It was observed that the needle was severely bent at the flash notch.The catheter was observed to be broken just distal of the luer.The distal segment of the catheter appeared bunched.The catheter was still attached to the needle, and the safety mechanism was not advanced over the needle tip.A microscopic observation of the catheter revealed a chevron shaped break with a granular fracture surface.The guidewire appeared to have misaligned coils along the distal tip of the guidewire.The distal weld tip of the guidewire was still intact.The observed characteristics of the break along with misaligned coils of the guidewire are typically caused by pulling the guidewire and catheter back against the needle bevel during the insertion process.The complaint of a broken catheter is confirmed and was determined to be use related.This complaint will be recorded for future trending and monitoring purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL KIT 22G X 8 CM WITH BIOPATCH AND PROBE COVER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
maddy vincent
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18773620
Report Number3006260740-2024-00763
Device Sequence Number15379060
Product Code FOZ
UDI-Device Identifier00801741140693
UDI-Public(01)00801741140693
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K162377
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberN/A
Device Catalogue NumberF322088PT
Device Lot NumberREHV2676
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2024
Supplement Date Received by Manufacturer05/06/2024
Initial Report FDA Received Date02/23/2024
Supplement Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-