| Model Number |
2002-7-50 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Could cause an interruption in therapy and hypoxia.
|
| |
|
Event or Problem Description
|
|
Can't keep tubing attached to oxygen bleed in stem for adding oxygen to circuits providing positive pressure.
|
| |
|
Event or Problem Description
|
|
Can't keep tubing attached to oxygen bleed in stem for adding oxygen to circuits providing positive pressure.
|
| |
|
Additional Manufacturer Narrative
|
|
Could cause an interruption in therapy and hypoxia.The issue described in the customer complaint most closely aligns with risk id "r5" a potential cause of hazard being " disconnect/unable to connect to o2 source " severity = 8/10, occurrence = 2, therefore rpn = 16 per rma-20017c rev 7-oxygen tubing and accessories product risk analysis.Given the complaint history review, there was one complaint about a similar issue reported in the 24 months that preceded the complaint report date.Per table 1a of the risk determination chart (ref-20001-c1 rev 1), a severity rating of 8/10 (serious) and an occurrence level of "improbable" results in a "low" risk with a "capa optional".Therefore, no capa is required.Reviewed sop-20021-c1 rev 1 in order to determine if an hhe was triggered.None of the triggers listed applied to this complaint.Therefore, an hhe is not required.
|
| |
|
Search Alerts/Recalls
|