On 16th february 2024 getinge became aware of an issue with one of our columns - 118001b4 - hybrid or table column, surface-mounted.As it was stated, during preparation of the patient for evar procedure, a table rotation issue occurred and the communication was lost.There were no reports of an error message on the hand controller.The patient was on the table and had already been anesthetized.The issue led to a delay of at least 30 minutes.The case was able to continue after manual reset of the table.There was no injury reported, however, we decided to report the issue based on the potential for serious injury if the situation, namely the delay in surgery resulting in prolonged anesthesia time, was to reoccur.
|
|
On 16th february 2024, getinge became aware of an issue with one of our columns - 118001b4 - hybrid or table column, surface-mounted.As it was stated, during preparation of the patient for evar procedure, a table rotation issue occurred and the communication was lost.There were no reports of an error message on the hand controller.The patient was on the table and had already been anesthetized.The issue led to a delay of at least 30 minutes.The case was able to continue after manual reset of the table.The incident was initially assessed as reportable due to a delay in surgery resulting in prolonged anesthesia time.Recently, the incident has been reevaluated as according to the medical expert's assessment, the procedural delay did not have any major clinical relevance in this situation.The scenario described in the record is considered as non-reportable.
|
|
Getinge became aware of an issue with one of our columns - 118001b4 - hybrid or table column, surface-mounted.As it was stated, during preparation of the patient for evar procedure, a table rotation issue occurred and the communication was lost.There were no reports of an error message on the hand controller.The patient was on the table and had already been anesthetized.The issue led to a delay of at least 30 minutes.The case was able to continue after manual reset of the table.The incident was initially assessed as reportable due to a delay in surgery resulting in prolonged anesthesia time.Recently, the incident has been reevaluated as according to the medical expert's assessment, the procedural delay did not have any major clinical relevance in this situation.The scenario described in the record is considered as non-reportable.It was established that the device failed to meet the manufacturer's specification due to intermittent communication errors caused by user error.The device was being used for patient treatment when the incident occurred.We currently do not have any information that would warrant further action towards the device manufacturing or devices on the market, however as per our complaint handling processes will continue to monitor the customer experiences with the device for any future information.The correction of b5 describe event or problem, d1 brand name, d4 catalog #, d4 serial #, d4 unique identifier (udi) #, h3a device evaluated by manufacturer, h3b device not eval provide code, h3c if other provide code ¿ explain, h4 device manufacture date and h6 medical device ¿ problem code fields deems required.This is based on the internal evaluation.Previous b5 describe event or problem: on 16th february 2024, getinge became aware of an issue with one of our columns - 118001b4 hybrid or table column, surface-mounted.As it was stated, during preparation of the patient for evar procedure, a table rotation issue occurred and the communication was lost.There were no reports of an error message on the hand controller.The patient was on the table and had already been anesthetized.The issue led to a delay of at least 30 minutes.The case was able to continue after manual reset of the table.There was no injury reported, however, we decided to report the issue based on the potential for serious injury if the situation, namely the delay in surgery resulting in prolonged anesthesia time, was to reoccur.Corrected b5 describe event or problem: on 16th february 2024 getinge became aware of an issue with one of our columns - 118001b4 hybrid or table column, surface-mounted.As it was stated, during preparation of the patient for evar procedure, a table rotation issue occurred and the communication was lost.There were no reports of an error message on the hand controller.The patient was on the table and had already been anesthetized.The issue led to a delay of at least 30 minutes.The case was able to continue after manual reset of the table.The incident was initially assessed as reportable due to a delay in surgery resulting in prolonged anesthesia time.Recently, the incident has been reevaluated as according to the medical expert's assessment, the procedural delay did not have any major clinical relevance in this situation.The scenario described in the record is considered as non-reportable.Previous d1 brand name: hybrid or table column, surface-mounted.Corrected d1 brand name: magnus operating table system.Previous d4 catalog#: 118001b4.Corrected d4 catalog#: n/a.Previous d4 serial#: (b)(6).Corrected d4 serial#: (b)(6).Previous d4 unique identifier (udi)#: n/a.Corrected d4 unique identifier (udi) #: (b)(4).Previous h3a device evaluated by manufacturer: no.Corrected h3a device evaluated by manufacturer: yes.Previous h3b device not eval provide code: other.Corrected h3b device not eval provide code: n/a.Previous h3c if other provide code - explain: device not returned to manufacturer.Corrected h3c if other provide code - explain: n/a.Previous h4 device manufacture date: 09/01/2023.Corrected h4 device manufacture date: 07/21/2023.Previous h6 medical device ¿ problem code: activation, positioning or separation problem|positioning problem||3009.Communication or transmission problem|||2896.Corrected h6 medical device ¿ problem code: communication or transmission problem|intermittent communication failure||4038.Use of device problem|||1670.
|