| Model Number |
LX3 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported that during surgery an ophthalmic microscope was not locked and moved.When doctor puts face up to lenses microscope rocked and bottom pivot.Procedure details and patient impact have not been provided.Additional information has been requested but none received till date.
|
| |
|
Additional Manufacturer Narrative
|
|
The servicing record (sr) review did not show record of the customer requesting service.No further information was able to be obtained from this customer.With no additional, related information provided, the reported event was not able to be confirmed.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A non-conformance-based review of the serial number was performed and a potential contributing factor to the reported complaint was identified.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|