Model Number 791-360
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a020503 captured the reportable event of seal compromised.
 
Event or Problem Description
It was reported to boston scientific corporation that a lithovue flexscope was used during a flexi ureteroscopy (urs) procedure performed on (b)(6) 2024.During preparation, the lithovue film was not attached when removed from the box and was assumed to be non-sterile.The procedure was completed with a second lithovue scope.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Event or Problem Description
It was reported to boston scientific corporation that a lithovue flexscope was used during a flexi ureteroscopy (urs) procedure performed on (b)(6) 2024.During preparation, the lithovue film was not attached when removed from the box and was assumed to be non-sterile.The procedure was completed with a second lithovue scope.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a020503 captured the reportable event of seal compromised.Block h10: the device was returned without the packaging.Analysis did not identify any defects with the returned device.During the functional inspection, the device had full articulation in both directions and the lithovue was tested for image on a lithovue touch pc.The device displayed a live, clear image.The reported complaint is not confirmed.Based on the most relevant information, the device met all manufacturing specifications and a risk review confirmed that the event is accounted for in the risk documentation.The reported complaint was not confirmed as the packaging was not returned.Based on all gathered information and the results of the product analysis, the complaint investigation conclusion code selected is cause not established, indicating, the investigation findings do not lead to a clear conclusion about the cause of the reported adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LITHOVUE
Common Device Name
URETEROSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18776903
Report Number2124215-2024-10878
Device Sequence Number4325161
Product Code FGB
UDI-Device Identifier08714729874812
UDI-Public08714729874812
Combination Product (Y/N)N
Initial Reporter CountryLO
PMA/510(K) Number
K153049
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/07/2025
Device Model Number791-360
Device Catalogue Number791-360
Device Lot Number0032387762
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/19/2024
Supplement Date Received by Manufacturer04/09/2024
Initial Report FDA Received Date02/26/2024
Supplement Report FDA Received Date05/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
Patient Weight93 KG
-
-