SYNTHES GMBH TIBIAL NAIL-ADVANCED / 10 360 / STERILE; NAIL, FIXATION, BONE Back to Search Results
Catalog Number 04.043.240S
Medical Device Problem Codes Device Slipped (1584); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D2b: additional product code: hwc.D9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E3: reporter is a j&j sales representative.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the sales rep informed about a patient that was involved in a case yesterday on (b)(6) 2024 around 10 pm.The hospital called and they were trying to put in a tna nail and as they put the faulty nail down and then couldn¿t get the ball tip out and they had to take the nail out completely and put the nail ex.The item number that they have the problem with is 04.043.240s and there was ball tip reaming rod they got the number which will be given all.It causes incredible delays and the sales rep doesn¿t know any other harm to the patient.It looks that the poly in the nail came loose and stuck on the ball tip reaming rod.The device will be returned.The sales rep also needs to have the company the replace the nail and ball tip reaming rod po charge.This report is for one (1) tibial nail-advanced / 10 360 / sterile.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: device history record (dhr) review conducted: a manufacturing record evaluation was performed for the finished device product code: 04.043.240s-us, lot no: 921p197.It was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 26-jul-2022, manufacturing site:jabil bettlach, expiry date:30-jun-2032.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL NAIL-ADVANCED / 10 360 / STERILE
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18777530
Report Number8030965-2024-02879
Device Sequence Number8383581
Product Code JDS
UDI-Device Identifier10886982297792
UDI-Public(01)10886982297792
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K201336
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.043.240S
Device Lot Number921P197
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2024
Supplement Date Received by Manufacturer03/05/2024
Initial Report FDA Received Date02/26/2024
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.5 REAMING ROD BALL TIP/950-STERILE
Patient SexUnknown
-
-