| Model Number |
799-EMVP-05-01 |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a 47-year-old male patient, the device stopped ventilating.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.An internal inspection resulted in no findings.The device was recertified and returned to the customer.Review of the device activity logs did not show any activity for the event date of 01/15/2024.No trend is associated with reports of this type.
|
| |
|