| Catalog Number |
72200873 |
| Medical Device Problem Codes |
Device Emits Odor (1425); Smoking (1585); Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that after a procedure, the dii controller had smoke and a burning smell.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case-(b)(4).The reported device was received for evaluation.A visual inspection revealed no physical damage.A functional evaluation was performed on the returned device and found a short circuit message in port a.Further investigation revealed a burnt resistor.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a component failure.Factors that could have contributed to the reported event include connecting a hand piece that is still wet from sterilization processing or connecting a shorted out hand piece causing damage to the motor controller pcb.No containment or corrective actions are recommended at this time.
|
| |
|