| Model Number |
TFAT00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
12/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient cant see with the lens and clinical reason for explant being - adaption to multifocal, loss in contrast sensitivity.The iol was explanted and exchanged for an unknown atiol following the initial implant procedure.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury/malfunction.No further reports required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information has been requested and received stating that there was a mechanical complication of the lens.Symptoms resolved.The reason for the event was patient non neuro adaptation to multifocal lens.Additional information received and reported that the iol was exchanged for the same lens model but two diopter more power indicating the correct lens power was not implanted initially.There is no reported defect or fault with the iol.
|
| |
|