ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX UV ABSORBING TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFAT00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 12/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient cant see with the lens and clinical reason for explant being - adaption to multifocal, loss in contrast sensitivity.The iol was explanted and exchanged for an unknown atiol following the initial implant procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury/malfunction.No further reports required.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been requested and received stating that there was a mechanical complication of the lens.Symptoms resolved.The reason for the event was patient non neuro adaptation to multifocal lens.Additional information received and reported that the iol was exchanged for the same lens model but two diopter more power indicating the correct lens power was not implanted initially.There is no reported defect or fault with the iol.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX UV ABSORBING TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18777886
Report Number1119421-2024-00345
Device Sequence Number8307932
Product Code MFK
UDI-Device Identifier00380652429757
UDI-Public00380652429757
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/22/2026
Device Model NumberTFAT00
Device Catalogue NumberTFAT00.220
Device Lot Number15204568
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2024
Supplement Date Received by Manufacturer02/20/2024
Initial Report FDA Received Date02/26/2024
Supplement Report FDA Received Date03/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON MONARCH IV IOL, INJECTOR; MONARCH III CARTRIDGE D; PROVISC
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-