| Catalog Number |
254500727 |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Attune ps fem trial sz 7 rt is cracked and chipped.Did the patient experience an adverse event such as infection, non-union, allergic reaction, osteoporosis, overloading, pain, degenerative diseases, bleeding or oozing? --> no, did the patient require revision surgery or hardware removal? --> unknown, patient status/ outcome / consequences --> no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown, (b)(4).Device property of -->none, device in possession of -->none, (b)(4).Device property of -->none, device in possession of -->none, (b)(4).Device property of -->none, device in possession of -->none, (b)(4).Device property of -->none, device in possession of -->none by checking this box i certify that all information that are known/available has been disclosed.If any new information will be made available, the additional information will be submitted through cst.--> true.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: according to the information received, ¿broken and cracked attune shim.Sz 5/6 12mm.Attune articulating surface sz8 missing spring.Attune femoral trial 6r chipped and sz 7r cracked and chipped¿.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that attune ps fem trial sz 7 rt exhibits deep scratches at the top section of the device.However, no cracks and signs of breakage were observed.No other defects that could have contributed to the observed condition of the trial were observed.This type of damage is consistent with with the device coming in contact with the saw blade while trialing, care should be taken to avoid saw blade excursion into the femoral trials or implants the overall complaint was confirmed as the observed condition of the attune ps fem trial sz 7 rt would have contributed to a device issue.Based on the investigation findings, the potential cause is traced to unintended user error, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|
|