| Model Number |
CD1257-40 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Failure to Interrogate (1332)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
|
| |
|
Event or Problem Description
|
|
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the implantable cardioverter defibrillator (icd) exhibited a failure to interrogate.The implantable cardioverter defibrillator (icd) was explanted.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported events were premature battery depletion and communication failure.The device was returned for analysis.Premature battery depletion and communication failure was confirmed by analysis.The loss of communication was due to low battery voltage; the low battery voltage was a result of premature battery depletion.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|
| |
|