W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; SYSTEM, ENDO GRAFT, ARTERIOVENOUS DIALYSIS ACCESS CIRCUIT STENOSIS TREATMENT Back to Search Results
Catalog Number VBHR090502A
Medical Device Problem Codes Break (1069); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2024, a patient underwent treatment for a thrombosed arteriovenous (av) fistula using a 9 mm x 5 cm gore® viabahn® endoprosthesis with heparin bioactive surface (viabahn® device) in the left axillary vein.It was reported the delivery catheter was advanced using an unknown brand / size introducer sheath over an unknown brand / size guidewire to the target treatment site.Reportedly, the deployment line was pulled and broke.The endoprosthesis did not deploy.It is unknown how the delivery catheter was withdrawn from the patient.It was reported that there was no impact to the patient.
 
Additional Manufacturer Narrative
A review of the manufacturing records indicated the device met pre-release specifications.Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Product return evaluation: engineering evaluation concluded the report of a broken deployment line to be consistent with the observations of a missing knob and lack of deployment line emerging from the hub during engineering evaluation.Device deployment continued with deployment line traction at the endoprosthesis indicating the zipper itself was not stuck.The root cause of the reported primary failure mode could not be established with the available information.Component code - g04044 - device deployer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE
Common Device Name
SYSTEM, ENDO GRAFT, ARTERIOVENOUS DIALYSIS ACCESS CIRCUIT STENOSIS TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
jorja nackard
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18779536
Report Number2017233-2024-04658
Device Sequence Number14532888
Product Code PFV
UDI-Device Identifier00733132645947
UDI-Public00733132645947
Combination Product (Y/N)Y
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P130006
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/17/2026
Device Catalogue NumberVBHR090502A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2024
Supplement Date Received by Manufacturer02/08/2024
Initial Report FDA Received Date02/26/2024
Supplement Report FDA Received Date04/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age22 YR
Patient SexMale
Patient Weight86 KG
-
-