| Catalog Number |
211816 |
| Medical Device Problem Code |
Self-Activation or Keying (1557)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 03/2026) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ultrasonic-guided biopsy procedure through normal density tissue, the physician turned the grip twice to set it, the inner needle allegedly shot out a little when it was set for second time and the device allegedly could not be able to collect tissue.No coaxial was used and the procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: received one monopty disposable core biopsy instrument for evaluation.Upon visual evaluation, the device appeared to have residue throughout the needle and the device was received in its initial position.No visual anomalies were noted.Upon functional testing, the device was able to be fully primed with no issues and the arrow was visible in the ready window.The device was able to prime and fire, however, during dry firing, the device self-activated.The device was disassembled, and the internal parts were inspected.No anomalies were noted to the hubs.Therefore, the investigation is confirmed for the reported self-activation issue as the device was self-activated during dry fire test.A definitive root cause for the reported self-activation could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 03/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ultrasound-guided biopsy procedure through normal density tissue, the physician turned the grip twice to set it, the inner needle allegedly shot out a little when it was set for second time and the device allegedly could not be able to collect tissue.No coaxial was used and the procedure was completed using another device.There was no reported patient injury.
|
| |
|