| Catalog Number |
05641489001 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The ca reagent lot number and expiration date were not provided.When the customer noticed the issue they attempted to recalibrate and calibrations failed.The customer stated qc was within range before the event.The field service engineer (fse) found an issue with a 3-way valve.The fse replaced the 3-way vacuum valve and verified the instrument by performing precision testing and qc.The service actions resolved the issue.
|
| |
|
Event or Problem Description
|
|
The initial reporter questioned low results for approximately 30 patient samples and multiple assays on a cobas 8000 c 701 module.Discrepant results were provided for 1 patient sample tested for calcium (ca).The initial result was 5.5 mg/dl.The repeat result from a different c 701 module was 9.5 mg/dl.The customer stated the questionable results for 1-2 patient samples may have been reported outside of the laboratory and corrected.The customer was not able to provide any clarifications on which patient samples were reported outside of the laboratory.The samples were repeated due to critical low results or delta checks.
|
| |
|
Search Alerts/Recalls
|