• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS 8000 C 701 MODULE; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS 8000 C 701 MODULE; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Catalog Number 05641489001
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The ca reagent lot number and expiration date were not provided.When the customer noticed the issue they attempted to recalibrate and calibrations failed.The customer stated qc was within range before the event.The field service engineer (fse) found an issue with a 3-way valve.The fse replaced the 3-way vacuum valve and verified the instrument by performing precision testing and qc.The service actions resolved the issue.
 
Event or Problem Description
The initial reporter questioned low results for approximately 30 patient samples and multiple assays on a cobas 8000 c 701 module.Discrepant results were provided for 1 patient sample tested for calcium (ca).The initial result was 5.5 mg/dl.The repeat result from a different c 701 module was 9.5 mg/dl.The customer stated the questionable results for 1-2 patient samples may have been reported outside of the laboratory and corrected.The customer was not able to provide any clarifications on which patient samples were reported outside of the laboratory.The samples were repeated due to critical low results or delta checks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 C 701 MODULE
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18787692
Report Number1823260-2024-00554
Device Sequence Number7757092
Product Code JJE
UDI-Device Identifier04015630928255
UDI-Public04015630928255
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K100853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05641489001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2024
Initial Report FDA Received Date02/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-