Model Number CNWTT0
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient was not able to read.Clinical reason as mentioned as astigmatism and subluxed iol exchange.The iol was exchanged for another lens model two months following the initial implant procedure.Additional information was requested.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the 20.0 diopter (add power +2.2/+3.2) lens was replaced with a 20.5 diopter (add power +2.2/+3.2) lens due to a report of astigmatism and that the iol subluxed.The product has not been received to evaluate.Follow up attempts were made.No further information has been provided at this time.File will be reopened when new information or the product is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC INTRAOCULAR LENS
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18788007
Report Number1119421-2024-00350
Device Sequence Number7814494
Product Code MFK
UDI-Device Identifier00380652408127
UDI-Public00380652408127
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNWTT0
Device Catalogue NumberCNWTT0.200
Device Lot Number15576407
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2024
Supplement Date Received by Manufacturer03/21/2024
Initial Report FDA Received Date02/27/2024
Supplement Report FDA Received Date04/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/30/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON PANOPTIX TRIFOCAL HYDROPHOBIC IOL
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-