| Model Number |
SIR-Y001 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The batch review and manufacturing records did not reveal any manufacturing issues with the sir-spheres.The device was not returned for investigation.This incident was initially assessed as a non-reportable event.A second review that was recently conducted, concluded that this event should have been reported.
|
| |
|
Event or Problem Description
|
|
Procedure included a segmental delivery with multiple aliquots.Everything prior and during the procedure looked normal.When the post procedure 4 point survey was done the healthcare professional realized that only 40% of the prescribed dose was delivered to the patient.No further information have been provided.
|
| |
|