| Model Number |
WA64360C |
| Medical Device Problem Codes |
Loose or Intermittent Connection (1371); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the root of the jaws was loose, and sparks appeared when output.The case was completed with the same set of equipment.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned.The evaluation found the insulation pad at the tip of the jaw was worn, and the insulation outer cover at the end was chipped.The wear of the insulation pad at the tip was caused by normal wear and tear during use, while the damage to the insulation outer cover at the end may be caused by the abrasion of the jaw core against the sleeve rod during installation or during cleaning and disinfection.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to field h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 years since the subject device was manufactured.The device was returned to olympus for inspection, and the customer's complaint that the root of the jaws was loose, and sparks appeared when output was confirmed.Based on the results of the investigation, it is likely that the event occurred due to the wear and tear and improper handling, frequent usage and reprocessing.Also, it can be assumed that severe hf flashover was caused by unintentional contact with other equipment or the distal jaws were constantly touching during hf application without tissue contact, resulting in a short circuit in the device.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|