It was reported to boston scientific corporation that a polaris loop ureteral stent was used during a flexible ureteral lithotripsy procedure in the kidney performed on (b)(6) 2023.During the procedure, it was noted that they had difficulty advancing the stent.The procedure was successfully completed with another polaris loop ureteral stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Investigation results revealed that the stent was torn, inside the patient, therefore this event has been deemed an mdr reportable event.Please see block h10 for full investigation details.
|
|
Block h6: imdrf device code a0414 captures the reportable investigation results of stent torn, inside the patient.Block h10: the returned polaris loop ureteral stent was analyzed, and a visual evaluation noted that there was no visual damage found.Additionally, the suture was not returned.During microscopic inspection, the renal coil was closely observed being torn.A functional inspection was performed and a mandrel of 0.039 was loaded into the device and no resistance was felt.No other problems with the device were noted.The reported event of stent difficult to advance was not confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.Based on the analysis performed, it may be related with some other factors such as excessive handling of the device, the technique used by the physician and normal procedural difficulties encountered during the procedure that could have possibly affected the device performance and its integrity.Therefore, all compiled information on this investigation determines that the most probable cause is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.
|