| Lot Number |
SFA00770 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that error 2 - 00000008 refrigerant flow obstruction detected occurred.During a pulmonary vein isolation (pvi) procedure, a smartfreeze cryoablation system console was selected for use.Approximately half-way through the procedure, after completion of ablations on the left side of the heart, the device was moved to the right side.During ablation, error 2 - 00000008 refrigerant flow obstruction detected occurred.The cryo gas cable connections were checked as well as replacement of the cable, catheter, and tank, however the issue persisted.The procedure was aborted prior to completion of ablations on the right side.There were no patient complications.The device will not be returned, as onsite repair was performed.Replacement of pv1 resolved the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory.The reported allegation is confirmed.The returned pressure valve was tested and it was confirmed that the valve was stuck closed as there was 0 flow and 0 injection pressure during the inflation state.When opening the valve, the plunger head was loose approximately 1 rotation.The 'manufacturing deficiency' code was selected for the reported allegation regarding "refrigerant flow obstruction".
|
| |
|
Event or Problem Description
|
|
It was reported that error 2 - 00000008 refrigerant flow obstruction detected occurred.During an ablation procedure a smartfreeze cryoablation system console was selected for use.After completion of ablations on the left side of the heart, the device was moved to the right side.During ablation, error 2 - 00000008 refrigerant flow obstruction detected occurred.The cryo cable, catheter, and tank were all replaced, however the issue persisted.The procedure was aborted prior to completion of ablations on the right side.There were no patient complications.The device is expected to be returned for analysis.It was further reported that the procedure was cancelled completely.
|
| |
|
Search Alerts/Recalls
|